A Comparative Study between EAR and Coaxial
Circle Breathing System
Theerapongpakdee S.,M.D.,Sermdamrongsak U.,M.D.
Department of Anesthesiology, Faculty of Medicine, Srinagarind
Hospital, Khon Kaen University, Khon Kaen 40002
Arterial blood gases were compared
in 20 patients of ASA class I-II between breathing system and enclosed afferent
reservoir (EAR) breathing system under general anesthesia with controlled ventilation
at 70 and 100 ml/kg/min FGF's.
The results indicated that at FGF of 100 ml/kg/min,there
were no statistically significant differences on pH,PaCO2
and PaO2,but
at FGF of 70 ml/kg/min there were statistically significant differences on pH and
PaCO2.
From this study we concluded that EAR breathing system was as efficient as coaxial
circle breathing system at FGF 100 ml/kg/min,but less efficient at FGF 70 ml/kg/min.
However,EAR circuit can be used safely at FGF 70 ml/kg/min because the results of
the study are within normal clinical range.