None
UWEDAT
Index
Glossary

Uwedat® Servicetool


Device Parameter

Introduction


Overview

Device Selector

Parametertable

Parameter Edit

Parametertable 1

Calculationtable

Parametertable 2

Parametertable 3

Parametertable 4

Parametertable 5


Menu

Main Menu Button

Help Button

Device Number

Device ON

Warm ON

Device OFF

Install

Delete

Load Button

Save Button

Load File Button

Save File Button

File Dialog


Parametertable 1

Image


Parameter

Component Number

Component Name

Serial Number

Signal Type

Device Code

Module Number (0-9)

Connector Number

Profile Number

Current Range 1-3

Range 1-3


Calculationtable

Image


Parameter


Parametertable 2

Image


Parameter

Tolerance

Signal Range

Conversion Par. 0-5

Conversion Factor

Units

Display Format

Measurement Cyclerate

Warm-UpTime


Parametertable 3

Image


Parameter

HMV Generation

HMV Availability 1, 2

Utility Component

Auto Messages

Auto Message Threshold

Auto Message Block-time


Parametertable 4

Image


Parameter

Calibration

Forced Calibration

Interval

HMV Exceedings

Threshold

Block-Time

Nominal Values

Tolerance Values

Prepare Time

Sample Time

Zerogas Cleansing

Recovery Time


Parametertable 5

Image


Parameter

Device Status Bits

Error Status Bits

HMV Status Bits


Uwedat® Servicetool

Device Parameter

Introduction

The environmental monitoring station is equipped with a number of measuring-devices (components) and each device is described by a set of parameters. The equipment and the parameterization is designated with the term configuration. The Device Parameter work-view is provided to manage the configuration. The device-menu includes controls to install and remove measuring-components, to manipulate parameter-sets and to start and stop the measurement of individual components. The device-parameter determines both the characteristics and behaviour of the measuring-components and the data conversion of the measured values.


Overview


The Device Parameter view is a tabbed pane divided into five editable tables. Each table contains a set of parameters, the tables are selected by click on the table-tab. If you want to view the parameter-set of a component, you must specify the component. You can enter the component-number directly or you can use the Device-Selector



Device Selector
The Device Selector is a external, floating window which is always on the top of the work-view window. The device-selector shows both the equipment (device-inventory) and the state of the individual devices (measuring, idle, edited). A device is selected via mouse-click, and unambiguous user-actions can be started with double-click.

The Device Selector is used with a couple of user-actions. For detailed information see Device Selector .







Parametertable
The parameters are grouped into three columns. The first column shows the meaning of the parameter and the second column shows the value of the parameter. The third column contains editable datafields which allows you to edit the parameter-value.

You can always view the parameter-set of individual components, but any changes to the parameter set are denied unless the measurement of the component is stopped (command Device Off).

Parameter Edit

The third column (titled Parameter New) contains editable entryfields which allows you to edit the parameter-value.
Entryfields are usually combined with formatters to prevent invalid data-input. If a formatter is applied with an entryfield, the used [formatting-rule] is described in rectangular brackets. It is important to know, that such entryfields doesn't accept data-input which violates the formatting-rules.

Parametertable 1

The parametertable 1 comprises common device- and signal-parameter. For detailed information see Parametertable 1.
Calculationtable

The calculation table gives the possibility to autocalculate the conversion parameters by the prior selected profile number, measuring range and manually edited device ranges. For detailed information see Calculationtable.
Parametertable 2

The parametertable 2 comprises parameters for signal conversion and signal tolerances. For detailed information see Parametertable 2.
Parametertable 3

The parametertable 3 comprises parameters for data evaluation (mean values) and data-qualifying. For detailed information see Parametertable 3.
Parametertable 4

The parametertable 4 comprises the description of individual bits of device-statusbyte, device-errorbyte and HMV -statusbyte. For detailed information see Parametertable 4.
Parametertable 5

The parametertable 5 comprises parameters for device calibration. For detailed information see Parametertable 5.



Menu


The Device Parameter Menu is used
to manage the equipment
(install and remove devices)
to change the state of devices
(measurement On, Off)
to load, manipulate and save device parameters
to copy device parameters to a backup-archive (harddisk or floppy)
All menu-actions requires the preceeding selection of one or multiple devices.


Main Menu ButtonQuit the Device-View and returns to the Main-View.
Help ButtonCalls the online-help for the Device-View (this chapter)
Device Number This entry-field is used to specify one single device directly by component-number. If you don't know the approbiate component-number, use the Device-Selector button to open the Device Selector . The Device Selector can also be used to make multiple selections at once.
Device ON Switches the measurement-state of the selected device to On (measuring). Multiple device-selections are allowed for this function.
Warm ON Switches the measurement-state of the selected device to On (measuring). No Startup-calibration and no warmup-procedure. Multiple device-selections are allowed for this function.
Device OFF Switches the measurement-state of the selected device to Off (idle). Multiple device-selections are allowed for this function.
Install Installs a new device. Requires a new unique component number. The install will be denied if the component-number already exists. This function is not applicable for multiple device-selections, you have to install devices step by step.

A usual approach to setup a new device is to load a prepared or a standard device-configuration from your backup-archive, set the proper device number and device-name whith the Servicetool and save the configuration to the station.

Delete Removes an existing device. The device must be "switched Off", otherwise the removing is denied. Multiple device-selections are allowed for this function.
Load Button Load the parameter-set of the selected device from the environmental monitoring station into the Servicetool and display them. Multiple device-selections are allowed for this function.
Save ButtonTransfer the (edited) parameter-set from the Servicetool back to the environmental monitoring station. Multiple device-selections are allowed for this function.
An usual approach to setup new devices is to duplicate the parameter-set from similar devices:
Install a new device with unique component-number
Load an existing similar device
Change the component-number in the parameter-table to match the new device
Select the new device with the Device Selector and save the parameter-set

Load File ButtonLoad a device configuration from a backup-archive (harddisk or floppy) to the Servicetool and display them. Use the file-dialog to select the approbiate file.

The backup-files for device-parameters usually have a filename which contains station-number and component-number "SnnnnKnnn.par" (german K for component!). Multiple devices usually stored with the filename "SnnnnKXXX.par".
EXAMPLES:
s1234k011.par = Component 011 from station 1234 s1234kxxx.par = Multiple components from station 1234

Save File ButtonSave the selected device-configuration to a backup-archive (harddisk or floppy). Use the file-dialog for subdirectory-selecting. A default-filename for the backup-file will be suggested (a combination of station-number and component-number with the file-extension "par").
File Dialog


Top of document

Parametertable 1

Image


Parameter

The parameters are grouped into three columns. The first column shows the meaning of the parameter and the second column shows the value of the parameter. The third column contains editable datafields which allows you to edit the parameter-value.

You can always view the parameter-set of individual components, but any changes to the parameter set are denied unless the measurement of the component is stopped (command Device Off).

Component Number Specifies the component type 01-99 and differentiates the same types with a subnumber 1-9.
LENGTH : 3 char
FORMAT : [Digit 0 - 9] >
EXAMPLE: 011

Component Name A short descriptive name of the component type.
LENGTH : 7 char
FORMAT : [ASCII 32 - 126]
EXAMPLE: Dust

Serial Number Serial or inventory number of the component.
LENGTH : 5 char
FORMAT : [ASCII 32 - 126]

Signal Type Defines how the measured value is acquired by the station computer. As an analog/digital value via the Intelligent Measuring Interface (IMI) or as a data string via a serial interface (SIO RS232).
R = RS232 serial interface
I = IMI Intelligent Measuring Interface
M = Meteodat datatelegram
K = IMKO soil moisture measurement
O = Opsis visual range measurement
S = SSM radioactivity measurement

Device Code Hexadecimal measuring device code, used with components that are connected via the measuring device interface (IMI). The device code is hardcoded in the cable-connector of the device.
LENGTH : 2 char
FORMAT : [Hexadecimal 0 - F]
EXAMPLE: 01

Module Number (0-9) Bus address of the interface module. Used with the measuring device interface IMI. The module-number is determined by the IMI frontpanel-switch.
LENGTH : 1 char
FORMAT : [Digit 0 - 9]

Connector Number Used with the measuring device interface (IMI) and the serial interface SIO. Possible numbers are IMI connector 1 - 3, SIO connector 1 - 6.
LENGTH : 1 char
FORMAT : [Digit 0 - 9]

Profile Number The profile number specifies what is measured (current, voltage, resistance etc.) with the measuring device interface (IMI). If the component is connected via serial interface (SIO), the profile number specifies the device-driver.
LENGTH : 1 char
FORMAT : [Digit 0 - 9]
EXAMPLE for IMI: see the measuring station user manual. EXAMPLE for SIO: 1 = NO, 2 = NO2, 3 = Dust, 4 = SO2, 5 = NOx

Current Range 1-3 Defines which of the three measurement ranges is used.
LENGTH : 1 char
FORMAT : [Digit 1 - 3]

Range 1-3 Specifies the measurement limits (in raw value units) for three different measurement ranges.
LENGTH : 12 char
FORMAT : [Digit 0 - 9] [E + -] noted in exponential format
EXAMPLE: from 0.00000E+00 to 1.00000E+03



Top of document

Calculationtable

Image


Parameter

The parameters are grouped into four columns. The first column shows the current profile number and their meaning, the second column shows the current measuring range. The third column contains the current conversion parameters. The fourth column contains editable datafields which allows you to edit the required device range and autocalculate the conversion parameters.

You can always view the parameter-set of individual components, but any changes to the parameter set are denied unless the measurement of the component is stopped (command Device Off).



Top of document

Parametertable 2

Image


Parameter

The parameters are grouped into three columns. The first column shows the meaning of the parameter and the second column shows the value of the parameter. The third column contains editable datafields which allows you to edit the parameter-value.

You can always view the parameter-set of individual components, but any changes to the parameter set are denied unless the measurement of the component is stopped (command Device Off).

Tolerance Limitation of the range in which the raw value is still valid, indicated in % of the selected measurement range.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] noted in %
EXAMPLE:
Upper Tolerance 05 % Lower Tolerance 03 %

Signal Range Range within which the measurement signal is valid and will be used for raw values generation.

ADC -4 to +4 Volt (-40 000 to 40 000)
Counter: 0 to 65 536
LENGTH : 12 char
FORMAT : [Digit 0 - 9] [E + -] noted in exponential format EXAMPLE:
Upper Signal Range 2.40000E+04
Lower Signal Range 2.00000E+02

Conversion Par. 0-5 Parameter (ki) for the polynomial for converting the measurement value (x) into the raw-value (y). For linear polynominals see Calculationtable.

Y=k5*x5+k4*x4+k3*x3+k2*x2+k1*x+k0
LENGTH : 12 char
FORMAT : [Digit 0 - 9] [E + -] noted in exponential format

Conversion Factor For converting the raw-value into a final-value wich has an alternative technical unit (e.g. wind velocity may be expressed in m/s or km/h or knots).
LENGTH : 12 char
FORMAT : [Digit 0 - 9] [E + -] noted in exponential format
EXAMPLE: 3.60000E+00

Units The unit of the raw value and the final value.
LENGTH : 10 char
FORMAT : [ASCII 32 - 126]
EXAMPLE: ppm

Display Format Format specification (decimal places before and after) for data output on screen and printer.
LENGTH : 2 char
FORMAT : [Digit 0 - 9]
EXAMPLES:
Format 40 1234
Format 31 123.4
Format 22 12.34
Format 13 1.234
Format 04 0.1234

Measurement Cyclerate Sampling rate for the measurement channel.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in seconds
EXAMPLE: 10 sec

Warm-UpTime Warm-up time in 10 minutes for components after start or power failure. During this time the final values are not used for averaging.
LENGTH : 3 char
FORMAT : [Digit 0 - 9] in minutes * 10
EXAMPLE: 003 (equals to 30 minutes)
Use Warm-Start to restart measuring-devices whitout this warm-up period.



Top of document

Parametertable 3

Image


Parameter

The parameters are grouped into three columns. The first column shows the meaning of the parameter and the second column shows the value of the parameter. The third column contains editable datafields which allows you to edit the parameter-value.

You can always view the parameter-set of individual components, but any changes to the parameter set are denied unless the measurement of the component is stopped (command Device Off).

HMV Generation Defines the type of data processing (see list below). Please note that with component groups all the components involved must use the same type of processing.
1 = arithmetic mean value
2 = summ
3 = peak value
4 = vector mean value
5 = Dust processing
6 = NOx processing
7 = windvector and gust
8 = windvector, gust with timestamp
9 = difference value, ie. NOx-NO
A = wind class distribution
B = triple

HMV Availability 1, 2 The HMV-Availability is determined by two parameters:
The availability 1 (lower limit - HMV invalid) defines the minimum percentage of good final values. Dropping under this value marks a HMV as "invalid".
The availability 2 (upper limit - HMV good) defines the percentage of good final values up to which a HMW is marked as "poor".
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in %
EXAMPLE : Availability 1 = 50%, 2 = 90%
Good values < 50% availability = invalid
Good values < 90% availability = poor
Good values > 90% availability = good

Utility Component Supplementary information for data generation in coded form, only required in special cases. Currently only used with wind speeds for definition of calm.
LENGTH : 3 char
FORMAT : [Digit 0 - 9]
000 = calm below 0.5 m/s
001 = calm below 2.0 km/h
002 = calm below 1 kn

Auto Messages Autogenerated messages (alarm) to the central monitoring system when the internal message threshold is exceeded. A scheduled shutdown of measuring-devices by means of relays can be initiated with this parameter.
J = Yes, generate auto message if threshold is exceeded
N = No, don't generate automessage
R = Relay, initiate scheduled shutdown
B = Both, auto message and scheduled shutdown

Auto Message Threshold If the measuring-value exceeds the threshold value, a error message is generated.
LENGTH : 12 char
FORMAT : [Digit 0 - 9] [E + -] in exponential format
EXAMPLE: 1.0000E+03, stated in final value units

Auto Message Block-time Suppresses successive internal messages.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in hours
EXAMPLE: 1 hour



Top of document

Parametertable 4

Image


Parameter

The parameters are grouped into three columns. The first column shows the meaning of the parameter and the second column shows the value of the parameter. The third column contains editable datafields which allows you to edit the parameter-value.

You can always view the parameter-set of individual components, but any changes to the parameter set are denied unless the measurement of the component is stopped (command Device Off).

Calibration The type of calibration is determined by four decisive parameters:
Zero-point Calibration?
Check-point Calibration?
Use calibration result?
Use invalid calibration result?
LENGTH : 1 char
FORMAT : [J = Yes][N = No]

Forced Calibration If a power failure exceeds the stipulated time period, a forced calibration is carried out (+warm-up).
LENGTH : 3 char
FORMAT : [Digit 0 - 9] in minutes
EXAMPLE: 030 minutes

Interval Determines the interval for periodic calibration. Normally daily, but <> 24 hours.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in hours
EXAMPLE: 23 hours

HMV Exceedings Number of HMV-Exceedings. This value determines how often the threshold-value may be exceeded before a forced calibration is carried out.
LENGTH : 2 char
FORMAT : [Digit 0 - 9]

Threshold Threshold value, stated in final-value units. If the accumulated number of threshold-exceedings reaches the limit given by HMV-exceedings, a forced calibration is carried out.
LENGTH : 12 char
FORMAT : [Digit 0 - 9] [E + -] in exponential format
EXAMPLE: 1.0000E+03, stated in final-value units

Block-Time Prevents multiple calibration within a certain time period.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in hours
EXAMPLE: 05 hour

Nominal Values Nominal values for zero- and check-point values, stated in raw-value units. If they are not reached (including tolerances), the calibration is marked as invalid.
LENGTH : 12 char
FORMAT : [Digit 0 - 9] [E + -] in exponential format
EXAMPLE: 1.0000E+03, stated in raw-value units

Tolerance Values Tolerances for calibration
Zero-point tolerance, in parts per thousand of measurement range (end value)
Check-point tolerance, in % of check-point nominal-value
LENGTH : 2 char
FORMAT : [Digit 0 - 9]

Prepare Time For rinsing the measuring chamber with zero or calibration gas.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in minutes

Sample Time During this time-period the zero-gas or calibration-gas is measured.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in minutes

Zerogas Cleansing Zero-gas after calibration-gas. For the final rinsing (cleaning) of the measuring chamber.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in minutes

Recovery Time Time-delay (idle period), before measurement recommences with sample air.
LENGTH : 2 char
FORMAT : [Digit 0 - 9] in minutes



Top of document

Parametertable 5

Image


Parameter

The parameters are grouped into three columns. The first column shows the meaning of the parameter and the second column shows the value of the parameter. The third column contains editable datafields which allows you to edit the parameter-value.

You can always view the parameter-set of individual components, but any changes to the parameter set are denied unless the measurement of the component is stopped (command Device Off).

Device Status Bits Meaning of the device status bits in plain text.
Error Status Bits Meaning of the device error bits in plain text.
HMV Status Bits Meaning of the HMV status bits in plain text. The meaning of the individual bits is defined by the station computer software (identical for all components) - please do not change the original text.


Top of document

1