I assume that the reader is familiar with the basic procedure of the
Foucault test. If not, I recommend that you read the chapters in
Texereaux's, Berry's or any other ATM book dealing with it.
Since this is the first Foucault tester I built, there are of course
many things I will do different next time. I also do not have direct
comparison to a normal slit-type tester as it is described in most of the
ATM books. I only intend to give one
example of many possible designs.
Testing with a slitless tester works basically the same way as it works
with a normal tester. For a general description of the testing procedure,
please refer to the common ATM literature.
It is generally difficult if not impossible to get perfect alignment of
the mirror and the tester, which means that you will have to shift the
tester sidewise often on to maintain the same slit size. One trick to
avoid this is to put pressure with your hand on the support where your
tester is placed. Unless your support is a heavy table this pressure will
make the tester tilt a little bit which provokes the same effect as
shifting the carriage sidewise.
Be patient and critical! A self made micrometer will surely introduce
some inaccuracy for which you have to account with at least 3-4 reading
sequences in each test set.
Another thing you have to avoid is to test in both directions at the
same time. You should in each test-sequence either turn the micrometer
clockwise or anticlockwise, which means to test either from zone 1 to 4
or from zone 4 to 1. Do not mix both movements.
Try NOT to look at the results of the
previous test sequences as you naturally will try to match previous
results. You should have independent readings in each set which will be
averaged to give a realistic result.
A great page about foucault testers!
This page has been visited times